This assay has high sensitivity and excellent specificity for detection of Eosinophil Chemotactic Factor (ECF).No significant cross-reactivity or interference between Eosinophil Chemotactic Factor (ECF) and analogues was observed.
Matrices listed below were spiked with certain level of recombinant Eosinophil Chemotactic Factor (ECF) and the recovery rates were calculated by comparing the measured value to the expected amount of Eosinophil Chemotactic Factor (ECF) in samples.
Matrix | Recovery range (%) | Average(%) |
serum(n=5) | 80-102 | 85 |
EDTA plasma(n=5) | 96-105 | 101 |
heparin plasma(n=5) | 96-105 | 101 |
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Eosinophil Chemotactic Factor (ECF) were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Eosinophil Chemotactic Factor (ECF) were tested on 3 different plates, 8 replicates in each plate. CV(%) = SD/meanX100 Intra-Assay: CV<10%>10%>Inter-Assay: CV<12%>12%>
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Eosinophil Chemotactic Factor (ECF) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample | 1:2 | 1:4 | 1:8 | 1:16 |
serum(n=5) | 79-88% | 95-102% | 97-105% | 93-101% |
EDTA plasma(n=5) | 87-98% | 99-105% | 93-101% | 83-104% |
heparin plasma(n=5) | 88-96% | 99-105% | 87-101% | 78-95% |
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
1. Prepare all reagents, samples and standards;2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;4. Aspirate and wash 3 times;5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;6. Aspirate and wash 5 times;7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;8. Add 50µL Stop Solution. Read at 450nm immediately.
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Eosinophil Chemotactic Factor (ECF). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Eosinophil Chemotactic Factor (ECF). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Eosinophil Chemotactic Factor (ECF), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Eosinophil Chemotactic Factor (ECF) in the samples is then determined by comparing the O.D. of the samples to the standard curve.
ELISA / CLIA Experiment Service
Single-component Reagents of Assay KitLysis Buffer Specific for ELISA / CLIAQuality Control of ELISA / CLIAELISA Kit Customized ServiceDisease Model Customized ServiceSerums Customized ServiceTGFB1 Activation ReagentReal Time PCR Experimental ServiceStreptavidin
Catalog No. | Organism species: Homo sapiens (Human) | Applications (RESEARCH USE ONLY!) |
APA025Hu01 | Active Eosinophil Chemotactic Factor (ECF) | Cell culture; Activity Assays. |
RPA025Hu01 | Recombinant Eosinophil Chemotactic Factor (ECF) | Positive Control; Immunogen; SDS-PAGE; WB. |
RPA025Hu02 | Recombinant Eosinophil Chemotactic Factor (ECF) | Positive Control; Immunogen; SDS-PAGE; WB. |
PAA025Hu01 | Polyclonal Antibody to Eosinophil Chemotactic Factor (ECF) | WB; IHC; ICC; IP. |
MAA025Hu22 | Monoclonal Antibody to Eosinophil Chemotactic Factor (ECF) | WB; IHC; ICC; IP. |
SEA025Hu | ELISA Kit for Eosinophil Chemotactic Factor (ECF) | Enzyme-linked immunosorbent assay for Antigen Detection. |
SCA025Hu | CLIA Kit for Eosinophil Chemotactic Factor (ECF) | Chemiluminescent immunoassay for Antigen Detection. |
在20mM Tris,150mM NaCl(pH8.0)中复溶至0.1-1.0 mg / mL的浓度。不要涡旋。
避免重复冷冻/解冻循环。在2-8°C下保存一个月。分装并在-80°C下存储12个月。
热稳定性由损失率描述。损失率通过加速热降解试验确定,即将蛋白质在37°C下孵育48h,没有观察到明显的降解和沉淀。在适当的存储条件下,有效期内的丢失率小于5%。